A direct comparison of liquid chromatography-mass spectrometry with clinical routine testing immunoassay methods for the detection and quantification of thyroid hormones in blood serum

Bowerbank, Samantha, Carlin, Michelle and Dean, John (2019) A direct comparison of liquid chromatography-mass spectrometry with clinical routine testing immunoassay methods for the detection and quantification of thyroid hormones in blood serum. Analytical and Bioanalytical Chemistry, 411 (13). pp. 2839-2853. ISSN 1618-2642

[img]
Preview
Text
Bowerbank2019_Article_ADirectComparisonOfLiquidChrom.pdf - Published Version
Available under License Creative Commons Attribution 4.0.

Download (794kB) | Preview
Official URL: https://doi.org/10.1007/s00216-019-01724-2

Abstract

A new and improved method was developed for the determination and quantification of four “free” thyroid hormones (i.e. 3,5-diiodothyronine (T2), 3,3′,5-triiodothyronine (T3), 3,3′,5′-triiodothyrone (rT3) and 3,5,3′,5′-tetraiodothyronine (T4)) in human serum by low- and high-resolution liquid chromatography-mass spectrometry (LC-MS). Several sample preparation strategies were investigated to obtain matrix-independent results. These strategies included solid phase extraction and matrix dilution. The developed analytical methods were then directly compared, in a blind study using patient-derived human blood serum samples, to the current clinical routine testing methods, i.e. electrochemiluminescence immunoassay and enzyme-linked immunosorbent assay. Chromatographic separation was achieved on a pentafluorophenyl (F5) column with an isocratic method of 30% aqueous phase, 70% organic phase where mobile phase A is 0.1% formic acid in water (pH 4) and mobile phase B is 0.1% formic acid in methanol (pH 4) (v/v). The high-resolution LC-MS was able to give a significant improvement in sensitivity with limits of quantification of 0.002 to 0.008 pmol/L for all four “free” thyroid hormones, as well as reduced sample preparation, making this the preferred method. However, the increase in capital cost may be beyond the capabilities of some laboratories. The LC-MS methods allow for the analysis of “free” thyroid hormones to be carried out in a significantly reduced analysis time. Clinical sample analysis showed that there was no statistical difference between the results obtained by ECLIA/ELISA and both LC-MS methods.

Item Type: Article
Uncontrolled Keywords: Liquid chromatography-mass spectrometry, Immunoassays, Thyroid hormones, Serum, Quantitative analysis
Subjects: F100 Chemistry
Department: Faculties > Health and Life Sciences > Applied Sciences
Depositing User: Becky Skoyles
Date Deposited: 31 May 2019 13:48
Last Modified: 01 Aug 2021 10:53
URI: http://nrl.northumbria.ac.uk/id/eprint/39432